Nanoparticle Analyzer SZ-100

Industry's widest range and highest precision measurement instrument for Nanoparticle characterization.

  • measures particle size, zeta potential, molecular weight and second virial coefficient;
  • particle diameter: 0,3 nm - 8,0 μm;
  • wide range of particle sizes and concentrations.

The SZ-100 nanopartica series instruments are flexible analytical tools for characterizing the physical properties of small particles. Depending on the configuration and application the system can be used as a particle size analyzer, or also used to measure zeta potential, molecular weight, (MW) and second virial coefficient (A2). Typical applications for the SZ-100 include nanoparticles, colloids, emulsions, and submicron suspensions.


Particle size analysis is performed by dynamic light scattering (DLS). Depending on the physical properties of the sample, the dynamic range is 0,3 nm – 8 µm. The lower limit is influenced by concentration, how strongly the sample scatters light, and the presence of large, unwanted particles. The upper limit is influenced by the density of the sample since DLS is modeled on all motion coming from Brownian motion, not gravitational settling.

The charge on the surface of particles is characterized by the SZ-100 by measuring the zeta potential of a suspension. The sample is injected into a disposable cell and a measurement of the particle electrophoretic mobility results in the calculated zeta potential. The zeta potential of the sample is most often used as an indicator of dispersion stability. Large magnitude zeta potential values indicate that an electrostatically stabilized suspension will remain stable. The zeta potential is often measured as a function of pH or other change in the chemistry to help formulators create new products with a long shelf life. Conversely identifying conditions at which the zeta potential is zero (that is, the sample is at the isoelectric point) allows one to choose optimum conditions for flocculating and separating particles.

The same instrument can also be used to measure the molecular weight and second virial coefficient of proteins, polymers, and other molecules. The user prepares several solutions with known concentrations and then uses the system in a static light scattering mode to create a Debye plot, which results in a calculation of both MW and A2.


Benefits:

  • Particle size, zeta potential, molecular weight, and second virial coefficient all in one instrument;
  • Wide range of particle sizes and concentrations;
  • Particles size measurements at both 90° and 173°;
  • Multiple particle size measurement modes for working with small particles and weak scatterers;
  • Small volume cells for both particle size and zeta potential.
Parameters Value

Particle size: Dynamic Light Scattering (DLS) technique

Measurement range

Particle diameter: 0.3 nm - 8.0 μm

Measurement Accuracy

Particle size: ISO 13321/22412 compliant
NIST traceable polystyrene latex particle standard: 100 nm measurement accuracy = +/- 2%

Measurement time

Approx. 2 minutes in general for particle size analysis

Sampling cell

Cuvette cell

Sampling volume

12 μl ~ 4 ml*
(* the sampling volume depends on the cell volume capacity.)

Zeta potential: Laser doppler electrophoresis technique

Measurement range

-200 ... +200 mV

Measurement time

Approx. 2 minutes in general

Sampling cell

Dedicated disposable cell or dip cell

Sampling volume

~100 μl for disposable cell

Molecular weight: Static light scattering Debye plot technique

Measurement Range

Mw: 1×103- 2 × 107 g/mol

Sampling cell

Cuvette cell

Physical

Power supply

AC 100-240 V, 50/60 Hz, 150 VA

Laser

DPSS 532 nm, 10 mW Class I

External dimensions

385(D) x 528(W) x 273(H) mm (except for projections)

Mass approx

25 kg